Weight |
1 kg |
Dimensions |
22 × 10 × 6 cm |
Electric safety:
- type of insulation: double, according to EN 61010 – 1 and IEC 61557
- measurement category: IV 600 V (III 1000 V) according to EN 61010 – 1
- protection class acc. to EN 60529: IP67
Other technical specifications:
- power suply: 4 alkaline batteries or or battery package Ni-MH
- weight: ~1 kg
- dimentions: 220 x 100 x 60 mm
Insulation resistance measurement
Measuring range according to EN 61557 – 2 for UN=50 V: 50 kΩ…250,0 MΩ
Range |
Resolution |
Accuracy |
0,0…999,9 kΩ |
0,1 kΩ |
±(3% m.v. + 8 digits) |
1,000…9,999 MΩ |
0,001 MΩ |
±(3% m.v. + 8 digits) |
10,00…99,99 MΩ |
0,01 MΩ |
±(3% m.v. + 8 digits) |
100,0…250,0 MΩ |
0,1 MΩ |
±(3% m.v. + 8 digits) |
Measuring range according to EN 61557 – 2 for UN=100 V: 100 kΩ…500,0 MΩ
Range |
Resolution |
Accuracy |
0,0…999,9 kΩ |
0,1 kΩ |
±(3% m.v. + 8 digits) |
1,000…9,999 MΩ |
0,001 MΩ |
±(3% m.v. + 8 digits) |
10,00…99,99 MΩ |
0,01 MΩ |
±(3% m.v. + 8 digits) |
100,0…500,0 MΩ |
0,1 MΩ |
±(3% m.v. + 8 digits) |
Measuring range according to EN 61557 – 2 for UN=250 V: 250 kΩ…2,000 GΩ
Range |
Resolution |
Accuracy |
0,0…999,9 kΩ |
0,1 kΩ |
±(3% m.v. + 8 digits) |
1,000…9,999 MΩ |
0,001 MΩ |
±(3% m.v. + 8 digits) |
10,00…99,99 MΩ |
0,01 MΩ |
±(3% m.v. + 8 digits) |
100,0…999,0 MΩ |
0,1 MΩ |
±(3% m.v. + 8 digits) |
1,000…2,000 GΩ |
0,001 GΩ |
±(3% m.v. + 8 digits) |
Measuring range according to PN-EN 61557 – 2 for UN=500 V: 500 kΩ…5,00 GΩ
Range |
Resolution |
Accuracy |
0,0…999,9 kΩ |
0,1 kΩ |
±(3% m.v. + 8 digits) |
1,000…9,999 MΩ |
0,001 MΩ |
±(3% m.v. + 8 digits) |
10,00…99,99 MΩ |
0,01 MΩ |
±(3% m.v. + 8 digits) |
100,0…999,0 MΩ |
0,1 MΩ |
±(3% m.v. + 8 digits) |
1,000…5,000 GΩ |
0,001 GΩ |
±(4% m.v. + 6 digits) |
Measuring range according to EN 61557 – 2 for UN=1000 V: 1000 kΩ…10,00 GΩ
Range |
Resolution |
Accuracy |
0,0…999,9 kΩ |
0,1 kΩ |
±(3% m.v. + 8 digits) |
1,000…9,999 MΩ |
0,001 MΩ |
±(3% m.v. + 8 digits) |
10,00…99,99 MΩ |
0,01 MΩ |
±(3% m.v. + 8 digits) |
100,0…999,0 MΩ |
0,1 MΩ |
±(3% m.v. + 8 digits) |
1,000…5,000 GΩ |
0,001 GΩ |
±(4% m.v. + 6 digits) |
5,00…10,00 GΩ |
0,01 GΩ |
±(4% m.v. + 6 digits) |
Continuity measurement of protective and equipotential conductors with the 200 mA current
Measuring range according to EN 61557 – 4: 0,10…1999 Ω
Range |
Resolution |
Accuracy |
0,00…19,99 Ω |
0,01 Ω |
±(2% m.v. + 3 digits) |
20,0…199,9 Ω |
0,1 Ω |
±(2% m.v. + 3 digits) |
200…999 Ω |
1 Ω |
±(4% m.v. + 3 digits) |
- Voltage on open terminals: <8 V
- Output current at R <2 Ω: ISC >200 mA: ISC >200 mA
- Compensation of test leads’ resistance
- Unidirectional current flow
Low-voltage and resistance measurement
Range |
Resolution |
Accuracy |
0,0…199,9 Ω |
0,1 Ω |
±(3% m.v. + 3 digits) |
200…1999 Ω |
1 Ω |
±(3% m.v. + 3 digits) |
- Voltage on open terminals: <8 V
- Current for closed terminals 5 mA < ISC < 15 mA
- Sound signal and green LED on when measured resistance < 30 Ω ± 50%
- Compensation of test leads’ resistance,
Capacitance measurements
Range wyświetlania |
Resolution |
Accuracy |
1…999 nF |
1 nF |
±(5% m.v. + 10 digits) |
1,00…9,99µF |
0,01µF |
±(5% m.v. + 10 digits) |
- Capacitance value displayed during the RISO measurement
- For test voltages below 100 V and measured resistance below 10 MΩ, unspecified capacitance measurement error
Measurement of alternating and direct voltage
Range |
Resolution |
Accuracy |
0,0…299,9 V |
0,1 V |
±(2% m.v. + 6 digits) |
300…600 V |
1 V |
±(2% m.v. + 2 digits) |
- Frequency range: 45…65 Hz
There are no reviews yet.