Key features

  • 600 V Cat III safety rating
  • 50% bigger display with bright white backlight
  • Voltage, resistance, continuity, capacitance
  • Input terminal for ac and dc current measurements to 10 A current
  • Diode test, data hold

21,058

Out of stock

Compare

Fluke quality is always within your reach

Your job requires that you have a rugged, reliable and accurate digital multimeter. The Fluke 15B+ digital multimeter does all that and more. Easy to use with one hand, even with gloves on, the Fluke 15B+ offers everything you need.

Category:

Specifications: Fluke 15B+ Digital Multimeter

Accuracy Specifications

Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0% to 75%. Accuracy specifications take the form of: ±(% of Reading + Number of Least Significant Digits).

Function Range Resolution Accuracy
AC volts
(40 Hz to 500Hz)1
4.000 V
40.00 V
400.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
1.0% + 3
DC volts 4.000 V
40.00 V
400.0 V
1000 V
0.001 V
0.01 V
0.1 V
1 V
0.5% + 3
AC millivolts 400.0 mV 0.1 mV 3.0% + 3
DC millivolts 400.0 mV 0.1 mV 1.0% + 10
Diode test2 2.000 V 0.001 V 10%
Resistance
(Ohms)
400.0 Ω
4.000 kΩ
40.00 kΩ
400.0 kΩ
4.000 MΩ
40.00 MΩ
0.1 Ω
0.001 kΩ
0.01 kΩ
0.1 kΩ
0.001 MΩ
0.01 MΩ
0.5% + 3
0.5% + 2
0.5% + 2
0.5% + 2
0.5% + 2
1.5% + 3
Capacitance3 40.00 nF
400.0 nF
4.000 μF
40.00 μF
400.0 μF
1000 μF
0.01 nF
0.1 nF
0.001 μF
0.01 μF
0.1 μF
1 μF
2% + 5
2% + 5
5% + 5
5% + 5
5% + 5
5% + 5
AC current μA
(40 Hz to 400 Hz)
400.0 μA
4000 μA
0.1 μA
1 μA
1.5% + 3
AC current mA
(40 Hz to 400 Hz)
40.00 mA
400.0 mA
0.01 mA
0.1 mA
1.5% + 3
AC current A
(40 Hz to 400 Hz)
4.000 A
10.00 A
0.001 A
0.01 A
1.5% + 3
DC current μA 400.0 μA
4000 μA
0.1 μA
1 μA
1.5% + 3
DC current mA 40.00 mA
400.0 mA
0.01 mA
0.1 mA
1.5% + 3
DC current A 4.000 A
10.00 A
0.001 A
0.01 A
1.5% + 3
Backlight Yes
1All ac, Hz, and duty cycle are specified from 1% to 100% of range. Inputs below 1% of range are not specified.
2Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA.
3Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 40 nF range).
Function Overload protection Input impedance (Nominal) Common mode rejection ratio Normal mode rejection ratio
AC volts 1000 V 1 >10 MΩ <100 pF >60 dB at dc,
50 Hz or 60 Hz
AC millivolts 400 mV >1MΩ, <100 pF >80 dB at 50 Hz or 60 Hz
DC volts 1000 V 1 >10 MΩ <100 pF >100 dB at dc,
50 Hz or 60 Hz
>60 dB at 50 Hz or 60 Hz
DC millivolts 400 mV >1MΩ, <100 pF >80 dB at 50 Hz or 60 Hz
1 106 V Hz Max
General specifications
Maximum voltage between any terminal and earth ground 1000 V
Display (LCD) 4000 counts, updates 3/sec
Battery type 2 AA, NEDA 15A, IEC LR6
Battery life 500 hours minimum (50 hours in LED Test mode without load. The hours with load depends on the type of LED under test.)
Temperature
Operating 0 °C to 40 °C
Storage -30 °C to 60 °C
Relative humidity
Operating humidity Non-condensing (<10°C)
≤90% RH at 10 °C to 30 °C;
≤75% RH at 30 °C to 40 °C
40 MΩ range
≤80% RH at 10 °C to 30 °C;
≤70% RH at 30 °C to 40 °C
Altitude
Operating 2000 m
Storage 12,000 m
Temperature coefficient 0.1 X (specified accuracy) /°C (<18 °C or >28 °C)
Fuse protection for current inputs 440 mA, 1000 V Fast Fuse, Fluke specified part only.
11A, 1000V Fast Fuse, Fluke specified part only.
Size (H x W x L) 183 x 91 x 49.5 mm
Weight 455 g
IP rating IP 40
Safety IEC 61010-1, IEC61010-2-030 CAT III 600 V, CAT II 1000 V, Pollution Degree 2
Electromagnetic environment IEC 61326-1: Portable
Electromagnetic compatibility Applies to use in Korea only
Class A equipment (industrial broadcasting & communication equipment)1
1 This product meets requirements for industrial (Class A) electromagnetic wave equipment and seller or
user should take notice of it. This equipment is intended for use in business environments and is not to be
used in homes.

Brand

FLUKE

Based on 0 reviews

0.0 overall
0
0
0
0
0

Be the first to review “Fluke 15B+ Digital Multimeter”

There are no reviews yet.

  • UT206A+

    0 out of 5
    (0)

    1. 1000A AC current measurement
    2. 1000V AC/DC voltage measurement
    3. 600A inrush current measurement (UT206A+)
    4. Temperature (UT206A+)
    5. NCV (UT206A+)
    6. ACV frequency 10Hz~10kHz
    7. Jaw opening: 45mm
    8. Resistance / capacitance / frequency measurement
    9. Auto range / backlight / low battery indication
    10. Continuity / diode / duty cycle
    11. MAX/MIN
    12. Data hold / flashlight / relative mode
    13. Auto power off
    14. CAT II 1000V/CAT III 600V

    SKU: n/a
  • UT60A Modern Digital Multimeters

    0 out of 5
    (0)

    Product Features

    1. Analogue bar graph

    2. Diode test and buzzer continuity

    3. Transistor measurement

    4. Auto range

    5. MAX/MIN/REL measurement mode

    SKU: 60AA075
  • LR1000 Laser Rangefinder

    0 out of 5
    (0)

    LR series laser rangefinders are portable long distance measurement tools. They are portable and ergonomically designed with high accuracy for a wide range of applications from leisure to professional including hunting, sailing, large scale hauling, large area surveying, and more. Its built-in rechargeable Lithium battery ensures long operation runtime. The 7x optical zoom design can measure distance, angle, and calculate Pythagorean relationships.

    SKU: UN100LR0T
  • LM120C Laser Distance Meters (Discontinued)

    0 out of 5
    (0)

    Features of LM80C/LM120C/LM200C Laser Distance Meters

    ● 2.4-inch color screen

    ● 12 measurement modes
    ● Electronic leveling
    ● 3 selectable measurement references
    ● USB 3.0 supply power
    ● 100 groups of data storage
    ● Rotatable screen
    ● Camera-aided measurement
    ● 2m drop test
    ● IP65 dust and water proof
    SKU: n/a
  • UT387C Wall Scanners

    0 out of 5
    (0)

    Features of UT387C/UT387D Wall Scanners

    Certificates: CE, UKCA, FCC

    ● Buzzer indication

    ● Auto power off

    ● Error calibration alarm

    ● LED alarm

    ● Low battery indication

    SKU: n/a
  • UT283A Single Phase Power Quality Analyzer

    0 out of 5
    (0)

    1. Trend/event capture

    2. Harmonics/inrush current measurement
    3. File management through USB interface
    4. Flick measure (PST/PLT values)
    5. User friendly interface
    6. Multimeter functions
    7. Scope functions including waveform display
    8. 3.5 inches 64K color TFT LCD display
    9. THD (Total Harmonic Distortion) function
    10.Long-term monitoring
    SKU: n/a