Insulation Resistance Meter
Insulation Tester
MIC-5001, 5000V Insulation Tester
Availability:
5 in stock
Insulation resistance measurement in a compact, light-weight, low-cost design.
- Measurement to 5 TΩ
- Measuring voltage selectable from 50 V to 500 V range in 50 V steps, and from 500 V to 5000 V range in 100 V steps
- Continuous indication of insulation resistance or leakage current
- Automatic discharge of stored charge due to capacitance of tested object after the insulation resistance measurement
- Audio beeps every five seconds to alert user test is proceeding, and to get time characteristics
- Measured test times of T1, T2 and T3 for measuring Absorption Coefficient, Dielectric Absorption Ratio DAR. and Polarization Index PI
- Continuous display of test voltage during the measurement
- Protection against measuring live objects
- Two and three-lead measurement method
- Ramp Test Insulation resistance measurement
- Measurement of breakdown voltage with increments up to 1 kV / s
- Measurement of alternating and direct voltages from 0 V to 750 V
- 990 cells of memory (11880 records) to record results
- Data transfer to PC via USB cable
- Built-in rechargeable battery pack
- Meets the requirements of the EN 61557 standard
- Charge from optional car lighter (12 V) socket accessory
₨0
5 in stock
Weight | 1 kg |
---|---|
Dimensions | 20 × 18 × 8 cm |
Insulation resistance measurement (two-lead)
Measurement range acc. to IEC 61557- 2 for RISOmin=VISOnom/IISOmax…5 TΩ (IISOmax=1 mA)
Range | Resolution | Accuracy |
---|---|---|
0.0 to 999.9 kΩ | 0.1 kΩ | ±(3% m.v. + 20 digits) |
1.000 to 9.999 MΩ | 0.001 MΩ | ±(3% m.v. + 20 digits) |
10.00 to 99.99 MΩ | 0.01 MΩ | ±(3% m.v. + 20 digits) |
100.0 to 999.9 MΩ | 0.1 MΩ | ±(3% m.v. + 20 digits) |
1.000 to 9.999 GΩ | 0.001 GΩ | ±(3% m.v. + 20 digits) |
10.00 to 99.99 GΩ | 0.01 GΩ | ±(3% m.v. + 20 digits) |
100.0 to 999.9 GΩ | 0.1 GΩ | ±(3% m.v. + 20 digits) |
1.000 to 5.000 TΩ | 1 GΩ | ±(4% m.v. + 50 digits) |
Values of measured resistance depending on measurement voltage
Voltage VISO | Measurement range |
---|---|
up to 100 V | 50 GΩ |
200 to 400 V | 100 GΩ |
500 to 900 V | 250 GΩ |
1000 to 2400 V | 500 GΩ |
2500 V | 2500 GΩ |
5000 V | 5 TΩ |
RampTest Insulation resistance measurement
Range | Resolution | Accuracy |
---|---|---|
0.0 to 999.9 kΩ | 0.1 kΩ | ±(5% m.v. + 40 digits) |
1.000 to 9.999 MΩ | 0.001 MΩ | ±(5% m.v. + 40 digits) |
10.00 to 99.99 MΩ | 0.01 MΩ | ±(5% m.v. + 40 digits) |
100.0 to 999.9 MΩ | 0.1 MΩ | ±(5% m.v. + 40 digits) |
1.000 to 9.999 GΩ | 0.001 GΩ | ±(5% m.v. + 40 digits) |
10.00 to 99.99 GΩ | 0.01 GΩ | ±(5% m.v. + 40 digits) |
100.0 to 999.9 GΩ | 0.1 GΩ | ±(5% m.v. + 40 digits) |
1.000 to 4.999 TΩ | 0.001 TΩ | ±(5% m.v. + 40 digits) |
The measurement of breakdown voltage in RampTest mode
Range | Resolution | Chosen UISO | Accuracy |
---|---|---|---|
25.0 V to 99.0 V | 0.1 V | <600 V | ± 5% m.v. ± 10 digits |
100 V to 600 V | 1 V | <600 V | ± 5% m.v. ± 4 digits |
25 V to 999 V | 1 V | >600 V | ± 5% m.v. ± 5 digits |
1.00 kV to 5.00 kV | 10V | >600 V | ± 5% m.v. ± 4 digits |
DC and AC voltage measurement
Range | Resolution | Accuracy |
---|---|---|
0 to 299.9 V | 0.1 V | ±(3% m.v. + 2 digits) |
300 to 750 V | 1 V | ±(3% m.v. + 2 digits) |
- frequency range: 45 to 65 Hz
VISO – measurement voltage
The acronym “m.v.” stands for a “measured reference value”
Electrical safety:
- type of insulation double, in acc. with EN 61010-1 and IEC 61557
- measurement category III 600 V (IV 300 V) in acc. EN 61010-1
- case protection rating in acc. EN 60529 IP6529: IP65
Other technical specifications::
- power supply of the meter NiMH 9.6 V – 2 Ah battery pack and power supply 12 V
- weight of the meter approx. 0.9 kg, 2 lbs
- dimensions 200 x 180 x 77 mm, 7.9 x 7 x 3 ins
- LCD segment display
- memory for measurement results: 990 cells, 11880 records
- transfer of measurement results data via USB
There are no reviews yet.