-8%

CMP-1006 Digital Clamp-on multimeter

Availability:

5 in stock


 

Additionally:

  • safe, insulated measuring clamps,
  • autoranging/manual range selection,
  • DATA HOLD function, for holding measured values,
  • backlit LCD,
  • DC ZERO mode of measurement for DC current, possibility to zero the display and reading relative actual value less stored zero value,
  • MAX/MIN function,
  • over range indication,
  • AUTO-OFF function.

31,587 34,221

5 in stock

Compare

(TRUE RMS) up to 1000 A.

Measurement of INRUSH current in start-up phase of electrical device (high frequency of sampling)

AC/DC voltage measurement (TRUE RMS) up to 600 V.

Resistance measurement and continuity test:

  • continuity test with acoustic signalling (beeper) for resistance below 40.

Temperature measurement (Fahrenheit and Celsius units).

Frequency measurement.

Duty cycle measurement.

Diode test.

Category:
Weight 0.3 kg
Dimensions 23 × 8 × 5 cm

Electric security:

  • measurement category: CAT III 600 V acc. to EN 61010 – 1:2004
  • protection class acc. to EN 60529: IP40

 

Other technical data:

  • power supply: 9 V battery type 6LR61
  • display: 6600 counts, backlit LCD
  • continuity test: threshold 40 Ω; test current < 0,5 mA
  • diode test: test current of 0,3 mA, typical open circuit voltage < 3 V DC, typical
  • low battery indication: BAT displayed
  • over range indication: 0L displayed
  • sampling rate: 2 times per second
  • INRUSH: integration time 100 ms
  • temperature sensor: type K thermocouple
  • input impedance: 10 MΩ (V DC and V AC)
  • AC bandwidth: 50…400 Hz (A AC and V AC)
  • auto power OFF: approx. 25 min.
  • dimensions: 229 x 80 x 49 mm
  • weight: 303 g
  • accordance with following standards: EN 61010 – 1:2004, EN 61010 – 2 – 032
  • quality standard: ISO 9001

 

Rated operational conditions:

  • internal clamp diameters: d1 = 36 mm/d2 = 52 mm
  • operating temperature: +5…+40°C
  • storage temperature: -20…+60°C
  • operating humidity: max 80% up to 31°C decreasing linearly to 50% at 40°C
  • storage humidity: <80%
  • operating altitude: max 2000 m

 

AC/DC current measurement (TRUE RMS)

Range Resolution Accuracy (DC) Accuracy (AC)
0…659,9 A 0,1 A ±(2,5% m.v. +5 dgt) ±(2,5% m.v + 8 dgt) for f = 50…60 Hz
660…1000 A 1 A ±(2,8% m.v. +8 dgt) ±(2,8% m.v. + 8 dgt) for f = 50…60 Hz

 

DC voltage measurement

Range Resolution Accuracy
0…6,599 V 0,001 V ±(1,5% m.v. + 3 dgt)
6,60…65,99 V 0,01 V ±(1,5% m.v. + 3 dgt)
66,0…600,0 V 0,1 V ±(1,5% m.v. + 3 dgt)

 

AC voltage measurement (TRUE RMS)

Range Resolution Accuracy
0…6,599 V 0,001 V ±(1,8% m.v. + 5 dgt) for f = 50…60 Hz
6,60…65,99 V 0,01 V ±(1,8% m.v. + 5 dgt) for f = 50…60 Hz
66,0…600,0 V 0,1 V ±(1,8% m.v. + 5 dgt) for f = 50…60 Hz

 

Resistance measurement

Range Resolution Accuracy
0,0…659,9 Ω 0,1 Ω ±(1,0% m.v. + 4 dgt)
0,660…6,599 kΩ 0,001 kΩ ±(1,5% m.v. + 2 dgt)
6,60…65,99 kΩ 0,01 kΩ ±(1,5% m.v. + 2 dgt)
66,0…659,9 kΩ 0,1 kΩ ±(1,5% m.v. + 2 dgt)
0,660…6,599 MΩ 0,001 MΩ ±(2,5% m.v. + 3 dgt)
6,60…66,00 MΩ 0,01 MΩ ±(3,5% m.v. + 5 dgt)

 

Frequency measurement

Range Resolution Accuracy
30,0…999,9 Hz 0,1 Hz ±(1.2% m.v. + 2 dgt) sensitivity:30…5 kHz:10 Vrms min. 5 kHz…15 kHz:40 Vrms min. for 20%…80% duty cycle
1,000…9,999 kHz 0,001 kHz ±(1.2% m.v. + 2 dgt) sensitivity:30…5 kHz:10 Vrms min. 5 kHz…15 kHz:40 Vrms min. for 20%…80% duty cycle
10,00…15,00 kHz 0,01 kHz ±(1.2% m.v. + 2 dgt) sensitivity:30…5 kHz:10 Vrms min. 5 kHz…15 kHz:40 Vrms min. for 20%…80% duty cycle

 

Temperature measurement

Range Resolution Accuracy
-20…760°C 0,1°C ±(3,0% m.v. + 5°C)
-4…1400°F 0,1°F ±(3,0% m.v. + 9°F)

 

Duty cycle measurement

Range Resolution Accuracy
10,0…94,9% 0,1% ±(1,2% m.v. + 2 dgt)
  • pulse width: 100μs…100 ms,
  • frequency range: 30 Hz…5 kHz,
  • sensitivity: 30 Hz…5 kHz: 10 Vrms, 5 kHz…15 kHz: 40 Vrms.

 

„m.v.”- measured value

 

Brand

SONEL

Based on 0 reviews

0.0 overall
0
0
0
0
0

Be the first to review “CMP-1006 Digital Clamp-on multimeter”

There are no reviews yet.

  • YH 133A ELEMENT FOR 907A,907C

    0 out of 5
    (0)

    Features Of Soldering Iron Heating Element YH-133A For 907A,907C:

    • Suitable for YIHUA 936, 937D, 878serial, 852 serial, 898D, 898BD, 862D, 8786D, 868D and so on.
    • If the soldering iron is burned, use the heating core to replace the old one.
    • Stainless steel materials.
    • Fast heating.
    • 100% brand new product.
    • Made of high-quality material.
    SKU: n/a
  • UT572 Advanced Earth Resistance Tester

    0 out of 5
    (0)
    UT572 is for measuring earth resistance and soil resistivity. It can automatically check interface connection and detect  interference  voltage/frequency, with data hold and indication functions. It can perform 2/3/4-wire measurements for different on-site  applications, such as power equipment wiring and lightning protection equipment.
    SKU: n/a
  • UT395A Professional Laser Distance Meter

    0 out of 5
    (0)

    UT395 series are professional laser distance meters with color LCD display, electronic multi-axis leveler and other special features. Its unique operating system, friendly user interface, and personalized settings option all make these devices the perfect choice for professional distance measurement.

    SKU: n/a
  • UT387B Wall Scanner

    0 out of 5
    (0)

    UT387 series wall scanners are great tools designed for construction and home improvement sector. These scanners can pick up hidden objects behind walls, such as metals, live electrical wires, and wood. UT387A is designed for dry wall, but it can also effectively scan hardwood flooring, plywood, and coated wooden walls. It has two scanning modes, these are used for scanning walls of different thickness.

    SKU: n/a